The individual contributions of pH and chloride concentration to the corrosion kinetics of bioabsorbable magnesium (Mg) alloys remain unresolved despite their significant roles as driving factors in Mg corrosion. This study demonstrates and quantifies hitherto unknown separate effects of pH and chloride content on the corrosion of Mg alloys pertinent to biomedical implant applications. The experimental setup designed for this purpose enables the quantification of the dependence of corrosion on pH and chloride concentration. The in vitro tests conclusively demonstrate that variations in chloride concentration, relevant to biomedical applications, have a negligible effect on corrosion kinetics. The findings identify pH as a critical factor in the corrosion of bioabsorbable Mg alloys. A variationally consistent phase-field model is developed for assessing the degradation of Mg alloys in biological fluids. The model accurately predicts the corrosion performance of Mg alloys observed during the experiments, including their dependence on pH and chloride concentration. The capability of the framework to account for mechano-chemical effects during corrosion is demonstrated in practical orthopedic applications considering bioabsorbable Mg alloy implants for bone fracture fixation and porous scaffolds for bone tissue engineering. The strategy has the potential to assess the in vitro and in vivo service life of bioabsorbable Mg-based biomedical devices.
Corrosion
,Corrosion kinetics
,Mechano-electrochemical coupled phase-field model
,pH effects